SPECIFICATIONS |
MODEL |
7305 |
7314 |
7315 |
7316 |
GROUND BOND |
Output AC Current |
3 - 30A |
1 - 40A |
1 - 30A |
1 - 60A |
Output AC Voltage |
6V, Maximum |
8V, Maximum |
6V, Maximum |
12V, Maximum |
Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
SETTINGS |
Max-Limit Range |
Resistance |
0-510mΩ |
0-120mΩ |
0-600mΩ |
0-200mΩ |
0-150mΩ |
0-510mΩ |
0-120mΩ |
0-600mΩ |
0-300mΩ |
0-150mΩ |
AC Current |
3.0-10.0A |
10.1-30.0A |
1.00-10.00A |
10.01-30.00A |
30.01-40.00A |
1.00-10.00A |
10.1-30.00A |
1.00-15.00A |
15.01-30.00A |
30.01-60.00A |
Resolution |
1mΩ |
Accuracy |
Same as Resistance Measurement Accuracy |
Min-Limit Range |
Resistance |
- |
0-600mΩ |
0-200mΩ |
0-150mΩ |
0-510mΩ |
0-120mΩ |
0-600mΩ |
0-300mΩ |
0-150mΩ |
AC Current |
- |
1.00-10.00A |
10.01-30.00A |
30.01-40.00A |
1.00-10.00A |
10.1-30.00A |
1.00-15.00A |
15.01-30.00A |
30.01-60.00A |
Resolution |
1mΩ |
Accuracy |
Same as Resistance Measurement Accuracy |
Lead Resistance Offset |
Range |
0 - 100 mΩ, Auto/manual |
Resolution |
1mΩ |
Accuracy |
±(2% of setting +2 counts) |
Dwell Time |
Range |
0, 0.5 - 999.9 sec (0 = continuous) |
Resolution |
0.1 sec |
Accuracy |
±(0.1% + 0.05sec) |
MEASUREMENT |
AC Current |
Range |
3.0 - 30.0A |
1.00 - 40.00A |
1.00 - 30.00A |
1.00 - 60.00A |
Resolution |
0.1A |
0.01A |
0.01A |
0.01A |
Accuracy |
±(2%of reading +1count) |
±(3%of reading +3counts) |
±(3%of reading +3counts) |
±(3%of reading +3counts) |
Resistance |
Range |
0-510mΩ |
0-120mΩ |
0-600 mΩ |
0-600 mΩ |
0-200 mΩ |
0-150 mΩ |
0-510 mΩ |
0-510 mΩ |
0-120 mΩ |
0-600 mΩ |
0-600 mΩ |
0-300 mΩ |
0-150 mΩ |
3.0-10.0A |
10.1-30.0A |
1.00-5.99 A |
6.00-10.00 A |
10.01-30.00 A |
30.01-40.00 A |
1.00-2.99 A |
3.00-10.00 A |
10.01-30.00 A |
1.00-5.99 A |
6.00-15.00 A |
15.01-30.00 A |
30.01-60.00 A |
Resolution |
1 mΩ |
Accuracy |
±(2%of reading +1count) |
±3% of reading +3 counts) |
±(2% of reading +2 counts) |
±(3% of reading +3counts) |
±(2% of reading +2 counts) |
±3% of reading +3 counts) |
±(2% of reading +2 counts) |
GENERAL |
Input Voltage AC |
115/230Vac±15%, 50/60Hz ± 5%, Fuse 6.3A for 7305/7315 model, 10A for 7314 model, 15A for 7316 model |
PLC Remote Control |
Input: Test , Reset,Withstand Processing ,(Memory1.2.3, Interlock for 7314/7315/7316 model only) |
Output: Pass , Fail , Processing , Reset Out , Start Out |
Memory |
5 memories |
10 memories, 3 steps/memory |
Display |
16×2 LCD with back light |
128×64 Graphic LCD |
Key Lock |
To prevent unauthorized alteration of the test Parameters |
Calibration |
Build-in software and external calibrated meters |
Verification |
- |
Build-in software driven verification menu to test fault detection circuits |
Alarm Volume Setting |
- |
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
Environment |
0-40℃, 20-80%RH |
Dimension |
280(W)×89(H)×370(D)mm |
215(W)×89(H)×430(D)mm |
215(W)×89(H)×370(D)mm |
430(W)×130(H)×400(D)mm |
Net Weight |
10Kg |
14Kg |
8.7Kg |
20Kg |
STANDARD ACCESSORIES |
Power Cord(10A) |
×1 |
×1 (15A) |
Fuses |
×2 (Including a spare contained in the fuse holder) |
×1(On Board) |
Interlock Disable Key(1505) |
- |
×1 |
×1 |
×1 |
Ground Bond Test Lead 30A ,1.5m(1103) |
- |
- |
×1 |
- |
Ground Bond Return Lead 30A ,1.5m(1104) |
- |
- |
×1 |
- |
Ground Bond Return Lead 30A ,1.6m(1122) |
×1 |
- |
- |
- |
Ground Bond Test Lead 30A,1.6m(1131) |
×1 |
- |
- |
- |
Ground Bond Test Lead 40A,1.6m(1137) |
- |
×1 |
- |
- |
Ground Bond Return Lead 40A,1.6m(1138) |
- |
×1 |
- |
- |
Ground Bond Test Lead 60A,1.65m (1160) |
- |
- |
- |
×1 |
Ground Bond Return Lead 60A,1.65m(1161) |
- |
- |
- |
×1 |